Mechanical response at peri-implant mandibular bone for variation of pore characteristics of implants: A Finite Element Study

Tytuł:
Mechanical response at peri-implant mandibular bone for variation of pore characteristics of implants: A Finite Element Study
Autorzy:
Sarkar, Sulagna
Sahu, Tikeshwar Prasad
Datta, Arijit
Chandra, Nimesh
Chakraborty, Arindam
Datta, Pallab
Majumder, Santanu
Chowdhury, Amit Roy
Tematy:
implant stomatologiczny
uszczelnianie
FEM
porous dental implant
pore size
percentage porosity
bone-implant interface
finite element analysis
stress and strain
Data publikacji:
2019
Wydawca:
Politechnika Wrocławska. Oficyna Wydawnicza Politechniki Wrocławskiej
Język:
angielski
Prawa:
Wszystkie prawa zastrzeżone. Swoboda użytkownika ograniczona do ustawowego zakresu dozwolonego użytku
Źródło:
Acta of Bioengineering and Biomechanics; 2019, 21, 2; 83-93
1509-409X
2450-6303
Dostawca treści:
Biblioteka Nauki
Artykuł
  Przejdź do źródła  Link otwiera się w nowym oknie
In this paper, the mechanical response of generic dental implants having calculated porosities with varying pore-sizes has been evaluated. The purpose of this study was to compare the developed stress-strain of designed porous implants (i.e., stress at the implant and strain at the peri-implant bone) with that of the non-porous implant. Methods: 3D model of a mandible was prepared from CT scan data and nine generic dental implant models have been designed having 10%, 20%, and 30% porosity with 500, 700, and 900 micron pore size along with a non-porous model for carrying out FE analyses. First, failure analyses of implants, under a biting force of 250 N have been performed. Next, the remaining implants have been further evaluated under average compressive chewing load of 100 N, for mechanical responses at bone-implant interface. Results: Von Mises strain at the peri-implant mandibular bone increases with the increase in percentage porosity of the implant material and maximum implant stress remained much below the yield stress level. Conclusion: Implant stiffness and compressive strength vary as a function of porosity and pore size. Strain obtained on the peri-implant bone is sufficient enough to facilitate better bone growth with the 700 micron pore size and 30% porosity, thus reducing the effect of stress shielding.

Ta witryna wykorzystuje pliki cookies do przechowywania informacji na Twoim komputerze. Pliki cookies stosujemy w celu świadczenia usług na najwyższym poziomie, w tym w sposób dostosowany do indywidualnych potrzeb. Korzystanie z witryny bez zmiany ustawień dotyczących cookies oznacza, że będą one zamieszczane w Twoim komputerze. W każdym momencie możesz dokonać zmiany ustawień dotyczących cookies